Case Study 5:  After Calibration and Re-Calibration, An Instrument Continues to Perform Nonlinear – all Levels are outside statistical limits and appear visually nonlinear

Initial Results:  A laboratory performed routine calibration verification / linearity testing using VALIDATE®TDM1 on a Roche Hitachi.  The following was the linearity report for Digoxin (DIGN) generated using MSDRx®, Maine Standards Company’s Data Reduction software:

Troubleshooting:  The results were not consistent with Peers or with typical product performance. The laboratory took the troubleshooting step of recalibrating their DIGN assay.  Calibration set points covered the method range of 0.3 – 5.0 ng/mL (calibration set points were 0, 0.5, 1, 2, 3 and 5 ng/mL).  After re-calibration, calibration verification / linearity testing was performed and results continued to be nonlinear.  The laboratory requested a service call from the instrument manufacturer.  During the service call, a probe alignment issue was discovered and corrected.  To confirm that the probe alignment issue was causing the nonlinear response, the laboratory re-ran the calibration verification / linearity testing.  The updated MSDRx® report shows that all Levels are within the statistical limits.

Summary:  As calibration is not intended to identify instrument issues, in this case, if the laboratory depended on calibration alone, the laboratory could have reported inaccurate patient results.  Calibration verification / linearity testing is the only way to test if a method is giving a nonlinear response.

Back to top

For Technical Support:

Phone:   800-377-9684
    207-892-1300
 
Fax:   207-892-2266

Contact us via email

 

Hours:
8:30 AM to 5:00 PM,
Eastern Time, Monday
through Thursday.
8:30 AM to 4:00 PM on Friday.

 

For technical consultation outside of normal business hours, simply call during business hours or email us to schedule a time.

FOR ORDERING, CUSTOMER SERVICE AND TECHNICAL SUPPORT, CALL 1-800-377-9684.