Initial Results: A laboratory performed routine calibration verification / linearity testing using VALIDATE® TDM1. Level 4 results for all of the analytes tested were higher than the target, causing statistical flags. The following was the report generated for Digoxin using MSDRx®, the Maine Standards Data Reduction software:
The laboratory contacted Maine Standards Company Technical Support. Technical Support advised the laboratory that the results were not consistent with Peers and that this curve was typical of product incompletely thawed and/or not thoroughly mixed. Frozen products need to be brought to room temperature and gently inverted several times before dispensing. Errors of this nature are most evident at higher concentrations.
Troubleshooting: The laboratory took the troubleshooting step of rerunning the calibration verification / linearity experiment for TDM1. The updated MSDRx® report shows that all Levels are within the statistical limits. The laboratory accepted the updated results and determined that they had validated the linearity across the reportable range of the method.
Summary: Maine Standards Company Technical Support advised that recovery was not consistent with Peers. In this case, inadequate thawing and/or mixing of the VALIDATE® product caused nonlinear response on the analyzer. Retesting of the material after properly thawing and mixing verified a linear response across the reportable range of the method.